Channel Sizes | Standard: 0.3, 0.5, 0.7, 1.0, 2.0, 5.0 μm |
Standard: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm |
Standard: 0.3, 0.5, 0.7, 1.0, 2.5, 5.0 μm |
Standard: 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm |
Optional: Other combinations (Please contact Us) |
Counting Efficiency | 50% at 0.3 μm; 100% for particles > 0.45 μm (per JIS) |
Concentration Limits | 4,000,000 particles / ft3 at 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count Level | |
Size Resolution | <15% at 0.5 μm (per ISO 21501-4) |
Flow Rate | 0.1 CFM (2.83 LPM) |
Flow Rate Control | Automatic flow control |
Calibration | NIST traceable |
Sample Probe/Tubing | Isokinetic sampling probe, Probe for tubing |
Sampling Modes | Manual, Automatic, ISO*1, GMP*2, Cumulative/Differential, |
Count/Concentration |
Reports | ISO 14644-1, EC GMP |
Sampling Time | 1 second to 99 minutes 59 seconds (Configurable) |
Sampling Frequency | 1 to 9999 cycles or continuous (Configurable) |
Sample Output | Internal HEPA filter |
Vacuum Source | Internal pump |
Communication Interface | USB |
Data Storage | 10000 sample records |
Security | 2-level password protection |
Alarm | Counts, Low battery, Flow, Laser |
Display | 3.5-inch 320 x 240 Color LCD |
Power | DC 5V 1A (Mini USB TYPE-B) |
Battery | 4 x AA |
Battery Life | Up to 4.5 hours of continuous use (LCD Backlight low, Included Ni-MH Battery) |
Environmental Sensors | Optional temperature/humidity probe |
Dimensions (L x W x H) | 178x90x47mm (without isokinetic inlet, temperature/humidity probe) |
Weight | 480g (without battery) |
Standards | CE, JISB9921, ISO 21501-4 |
Warranty | 2 year limited warranty |
Operating Conditions | 5° to 35°C 20% to 95%RH non-condensing |
Storage Conditions | -20° to 50°C Up to 98%RH non-condensing |
Included Accessories | Operating manual on CD, Quick guide, AC adapter, Isokinetic inlet, Probe for tubing, USB cable, Zero filter, Software, 4 x AA batteries with charger, Calibration certificate, Carrying case |
Optional Accessories | Temperature/Humidity probe |